PFAS‑Free Heat Transfer Fluids in 2026: What Labs Should Know Before Draining the Bath

Regulatory Pressure: Why "+ PFAS‑Free Heat Transfer Fluid" Matters for Labs in 2026

By late 2025 and into 2026, the regulatory landscape for heat transfer fluids is being reshaped on both sides of the Atlantic. Laboratories using per- and polyfluoroalkyl substance (PFAS)-based fluids—especially perfluoropolyether (PFPE)—now face growing compliance and operational risks.

What’s Fueling the Shift?

  • EU Action: The European Chemicals Agency (ECHA) will deliver its final sector opinions on PFAS restrictions by late 2025, with broad restrictions on PFAS—including most PFPEs—expected in 2026. While some applications (such as precision equipment and critical infrastructure) might get temporary sector-specific derogations, generic lab and industrial use are likely to be impacted most severely. Labs across the EU must prepare for emission monitoring, reporting mandates, and rapid compliance timeframes. (source)
  • US Action: Under the Toxic Substances Control Act (TSCA), facilities must now report PFAS use and emissions, and the EPA signals that phaseouts or additional restrictions on PFPE are probable after 2026. Many states already restrict PFAS in new chemistries and supply chains.

Early Takeaway

If your facility relies on PFPE for -80°C freezers, bath circulators, or process heating, now is the strategic window to evaluate alternatives. Late preparation may risk operational delays, unplanned downtime, or even insurance issues due to noncompliance.


PFAS-Free Heat Transfer Fluids: Technical Realities

Key Performance Factors

  • Flash Point & Flammability:PFPE fluids (like Fomblin®/Galden®) have no flash point, making them nearly non-flammable and preferred for operations where ignition risk is a top concern. In contrast, PFAS-free options—primarily silicone-based (PDMS) or engineered glycols—feature lower, defined flash points and are classified as combustible under NFPA 30.

  • Viscosity & Pump Load:PFPEs provide stable, moderate viscosities even at ultra-low temps, enabling smooth recirculation. PFAS-free fluids can be more viscous at subzero or high temps, potentially requiring stronger pumps and resulting in higher operational power draw. Always review your circulator’s pump curve and heater derate before switching.

  • Thermal Capacity:PFPEs generally offer higher heat capacities and maintain temperature setpoint stability across wide windows. Silicone and glycol fluids are slightly less efficient, which means more frequent cycling, longer ramp times, and potentially broader temperature fluctuations at the process.

  • Material & Seal Compatibility:Not all elastomers and plastics tolerate alternative fluids well. For example:

  • EPDM and FKM (Viton®) are often compatible with silicone and glycol fluids, while EPDM performs best at lower temperatures.

  • VMQ is optimal for silicones but may have limits above 200°C.

  • FFKM seals (perfluoroelastomer) are extremely heat-resistant but may still contain PFAS fractions and are costly.Always examine your system’s o-rings, hoses, and gaskets. Incompatibility risks leaks, swelling, or embrittlement—leading to downtime or equipment damage. (compatibility chart)

  • Dielectric Strength & Conductivity:If your application is sensitive to electrical leakage (e.g., high-voltage baths for semiconductor synthesis), silicone oils may perform as well as PFPE, but glycols may not. Check your safety and process specs closely.


Regulatory Compliance & Facility Risk Management

Fire Code and Insurance—New Priorities

Switching from a non-flammable PFPE bath to a flammable or combustible PFAS-free fluid triggers a new risk profile:

  • NFPA 30 and Factory Mutual (FM) codes govern storage, signage, and handling of flammable and combustible liquids. Your local Authority Having Jurisdiction (AHJ) likely enforces these provisions. (see NFPA 30 overview)
  • Insurance reviews may require you to update fire risk assessments, hazard signage, and, in some cases, sprinkler or spill mitigation measures before shipment and commissioning of new fluids.

Tip: Always notify your risk manager and AHJ before fluid changeover, and ensure the new SDS is on hand—today’s suppliers are publishing more thorough data sheets as a sign of compliance leadership.


Best Practices: Drain, Clean, Validate—A Stepwise Switchover

Changing a heat transfer bath fluid is not a simple drain-and-fill; it’s a precise operation requiring planning and documentation:

1. Schedule System Repairs & Assessments

  • Coordinate repairs and preventive maintenance with the switchover period to save downtime costs. Inspect for degraded seals or hoses, and document pump, heater, and safety sensor functional checks (changeout guide).

2. Drain the System

  • Turn off and cool the bath (typically for 24 hours). Circulate existing fluid to loosen deposits, then drain thoroughly to avoid mixing.

3. Clean and Rinse

  • Use heated compatible cleaning fluid if recommended by the new fluid supplier. Flush the system repeatedly until particulates and residues are below detection. Some vendors supply fluid test kits to check residue and compatibility (Eastman guide).

4. Inspect and Replace Seals & Hoses

  • Swap out incompatible gaskets, o-rings, and flexible hoses as required. Reference the elastomer compatibility findings above.

5. Refill, Commission, and Sample

  • Add PFAS-free heat transfer fluid slowly while monitoring for leaks and sensor faults. Run a gradual setpoint ramp and monitor system thermals for stability. Sample the fluid after 24–48 hours and send for analysis if required by your QA/QC protocols.

6. Update SOPs and Train Your Team

  • Document new handling, spill, and emergency procedures for the incoming fluid. Update SDS binders and alarm response plans.

Urth & Fyre Value: Beyond the Fluid

Choosing a compliant PFAS-free heat transfer fluid isn’t just about the liquid—it’s about facility safety, lifetime total cost, and assurance that every part of your system (from hose to SCADA interface) will perform as intended.

At Urth & Fyre, we help:

  • Evaluate your target temperature window and match it with a fluid’s performance.
  • Assess material compatibility and downstream maintenance scheduling.
  • Coordinate commissioning, initial fluid sampling, and validation workflows.
  • Support updates to your FM/NFPA documentation and insurance requirements.
  • Provide bath circulators and expertise to optimize thermal management—such as the Julabo SL-12 300°C 12L Heating Circulator, engineered for modern labs needing flexibility across legacy and next-generation fluids.

Common Pitfalls to Avoid

  • Skipping pump curve/power draw reviews after switching to more viscous fluids.
  • Overlooking elastomer and hose swapouts, exposing the system to leakage or catastrophic failure.
  • Failing to update AHJ or insurance records—risking liability or denial of coverage.
  • Insufficient training/documentation for lab operators handling flammable heat transfer fluids.

Key Takeaways for Lab Managers and Facilities Teams

  • Regulatory compliance for PFAS will be enforced rapidly from late 2025—laboratory service windows for switchover are shrinking.
  • Thermal and safety performance are non-negotiable. New fluids must be carefully matched, tested, and validated—not just swapped.
  • Total system review—from seals to risk documentation—prevents costly downstream failures.
  • Trust partners who can support not just equipment, but also workflow, compliance, and maintenance roadmaps.

If you’re facing the heat transfer fluid transition, contact us for lab equipment, consulting, and custom commissioning. Visit Urth & Fyre to explore fluid-ready bath circulators like the Julabo SL-12, and line up your compliance and safety teams ahead of the 2026 PFAS crunch.

Keywords: PFAS-free heat transfer fluid, PFPE, bath circulator, silicone glycol, NFPA 30, elastomer compatibility, lab compliance, material compatibility, pump curve, fire risk, commissioning, maintenance, ECHA, TSCA, safety audit, Julabo SL-12.